Summary
Type: TLC
Application #: 403540
Catalog#: 3415131 - Silica Chiral TLC Plates, w/UV254, glass backed, 250um, 10x20cm 4/pk
Detection:
Conditions: Plate pretreatment: plates were washed with methanol (1 min); after evaporation of solvent they were activated in an oven (105 °C, 15 min); plates were cut into 2 x 10 cm pieces
TLC
Applied volume: 2.5 µL
Eluent: acetonitrile – methanol – bidistilled water – 5 mM sodium acetate buffer + acetic acid, pH 5.57 (70:10:10:10, v/v/v/v)
Development: horizontal TLC chamber, equilibration time 15 min with eluent vapor, development time 4.5 min
PPEC (pressurized planar electrochromatography)
Applied volume: 2.5 µL on prewetted plates
Eluent: acetonitrile – methanol – bidistilled water – 5 mM sodium acetate buffer + acetic acid, pH 5.57 (70:10:10:10, v/v/v/v)
Polarization potential: 1.7 kV
Development: PPEC chamber (adapted from horizontal TLC chamber) with a high-voltage power supply (max. 2.5 kV), equilibration time 15 min with eluent vapor, development time 7.5 min
Authors: Polak, B. et al.
Source: Journal of Planar Chromatography 21 (2008) 1, 33–3